FluoroMax-3 v. 3.1 (7 Mar 2006)
Components & Accessories
12-32
Note:
The plastic cap also prevents
light from entering the sample
compartment.
3
Install purge port.
a
Remove plastic original plug, using
a 3/16" Allen key.
b
Screw in manually the new metal purge port.
c
Tighten with a 7/16" wrench.
d
Remove the plastic cap whenever a
hose is to be attached to the new purge
port.
4
Close the instrument’s cover.
New purge port.