FluoroMax-3 v. 3.1 (3 Mar 2006)
Maintenance
6-6
Note:
Do not replace
the FluoroMax
®
-3
cover until the lamp
is correctly adjusted.
lamp to the approximate position of the old lamp.
11
Replace the lamp-housing cover.
12
Replace the two screws on the lamp-housing
cover.
13
Reconnect all cables
(power, accessories, etc.)
to the FluoroMax
®
-3.
Adjusting the new xenon lamp
Choices
There are two choices after lamp installation:
•
Let the lamp “burn in”, i.e., run, for 6 h before adjustment of its position.
•
Set the coarse lamp adjustments immediately. After a 6-h burn-in, set the fine ad-
justments.
For lamp replacement, the major adjustment is to optimize the height screw that was
loosened to remove the old xenon lamp.
Procedure
1
With the instrument
cover still removed,
turn on the Fluoro-
Max
®
-3.
2
Insert a clean cuvette filled with de-ionized wa-
ter in the sample compartment.
3
With the room lights off and ambient stray light
minimized, acquire a xenon-lamp scan and wa-
ter Raman scan, as explained in Chapter 3.
Caution:
Intense ultraviolet, visible, or infrared light
may be present when the sample compartment is
open, so wear eye- and skin-protection, such as
light-blocking goggles and light-blocking clothing.
Содержание FluoroMax-3
Страница 5: ...FluoroMax 3 v 3 1 3 Mar 2006 v 15 Declaration of Conformity 15 1 16 Index 16 1...
Страница 6: ...FluoroMax 3 v 3 1 3 Mar 2006 vi...
Страница 18: ...FluoroMax 3 v 3 1 3 Mar 2006 Introduction 0 12...
Страница 36: ...FluoroMax 3 v 3 22 Feb 2005 System Description 2 8...
Страница 96: ...FluoroMax 3 v 3 1 3 Mar 2006 Maintenance 6 12...
Страница 156: ...FluoroMax 3 v 3 1 6 Mar 2006 Automated Polarizers 10 22...
Страница 208: ...FluoroMax 3 v 3 1 6 Mar 2006 Bibliography 14 4...
Страница 210: ...FluoroMax 3 v 3 1 6 Mar 2006 Declaration of Conformity 15 2...
Страница 220: ...FluoroMax 3 v 3 1 25 Apr 2006 Index 16 10...