FluoroMax-3 v. 3.1 (6 Mar 2006)
Automated Polarizers
10-9
Note:
Do not
check the Re-
m
easure Anisot-
ropy Only
check-
box.
Warning:
Read the Materials Safety
Data Sheets (MSDS) before using
colloidal silica or glycogen.
Choose various options:
•
Subtract
Dark
(recommended)
•
Reset to Mechanical Zero
—only if the polarizers
are definitely miscalibrated. This deletes the
previous calibration.
6
Place the LUDOX
®
or
glycogen in the sample
holder.
7
Click
the
Con-
tinue button.
The system rotates through the polarizers as shown in the checklist on the win-
dow. As each phase is completed, the checkboxes are updated.
If the sample is too
concentrated or dilute, the
software prompts you to
correct this.
Содержание FluoroMax-3
Страница 5: ...FluoroMax 3 v 3 1 3 Mar 2006 v 15 Declaration of Conformity 15 1 16 Index 16 1...
Страница 6: ...FluoroMax 3 v 3 1 3 Mar 2006 vi...
Страница 18: ...FluoroMax 3 v 3 1 3 Mar 2006 Introduction 0 12...
Страница 36: ...FluoroMax 3 v 3 22 Feb 2005 System Description 2 8...
Страница 96: ...FluoroMax 3 v 3 1 3 Mar 2006 Maintenance 6 12...
Страница 156: ...FluoroMax 3 v 3 1 6 Mar 2006 Automated Polarizers 10 22...
Страница 208: ...FluoroMax 3 v 3 1 6 Mar 2006 Bibliography 14 4...
Страница 210: ...FluoroMax 3 v 3 1 6 Mar 2006 Declaration of Conformity 15 2...
Страница 220: ...FluoroMax 3 v 3 1 25 Apr 2006 Index 16 10...