FluoroMax-3 v. 3.1 (6 Mar 2006)
FluoroMax
®
-P Phosphorimeter Operation
9-13
b
Gently slide the sample mount out of the instrument.
Some sample mounts have a 15-pin connector at the inside end for automated
accessories.
4
Remove the instrument cover.
a
Remove the seven Phillips screws (two on each side, and three in the
back) from the cover.
b
Lift the cover vertically off the instrument by grasping opposite corners
of the cover and raising upward.
If the cover sticks, gently work each side upward until the cover slides smoothly
off.
5
Remove the lamp-housing cover at the rear of
the instrument.
a
Remove the two screws from the
lamp-housing cover.
b
Gently rotate the lamp-housing
cover with its cooling fans at-
tached.
c
Rotate the cover backwards, and
set it behind the instrument so
that electrical connections are
not strained.
6
Remove the flash lamp.
a
Follow all safety precautions on the new lamp’s
box.
b
Pull the old flash lamp out with a steady motion.
Содержание FluoroMax-3
Страница 5: ...FluoroMax 3 v 3 1 3 Mar 2006 v 15 Declaration of Conformity 15 1 16 Index 16 1...
Страница 6: ...FluoroMax 3 v 3 1 3 Mar 2006 vi...
Страница 18: ...FluoroMax 3 v 3 1 3 Mar 2006 Introduction 0 12...
Страница 36: ...FluoroMax 3 v 3 22 Feb 2005 System Description 2 8...
Страница 96: ...FluoroMax 3 v 3 1 3 Mar 2006 Maintenance 6 12...
Страница 156: ...FluoroMax 3 v 3 1 6 Mar 2006 Automated Polarizers 10 22...
Страница 208: ...FluoroMax 3 v 3 1 6 Mar 2006 Bibliography 14 4...
Страница 210: ...FluoroMax 3 v 3 1 6 Mar 2006 Declaration of Conformity 15 2...
Страница 220: ...FluoroMax 3 v 3 1 25 Apr 2006 Index 16 10...