FluoroMax-3 v. 3.1 (3 Mar 2006)
System Operation
3-13
Note:
Observed throughput (and hence peak intensity) is af-
fected by lamp age and alignment, slit settings, and sample pu-
rity. As the xenon lamp ages, the throughput of the system will
decline slowly. Therefore, low water-Raman peak intensity may
indicate a need to replace the xenon lamp.
8
If the water-Raman peak is not at 397 nm, cali-
brate the emission monochromator as shown for
the excitation-monochromator calibration on
pages 3-7 to 3-9.