FluoroMax-3 v. 3.1 (3 Mar 2006)
Optimizing Data
5-6
Measuring the
G
factor
Include the grating factor, or
G
factor, whenever polarization measurements are taken.
The
G
factor corrects for variations in polarization wavelength-response for the emis-
sion optics and detectors. A pre-calculated
G
factor may be used when all other ex-
perimental parameters are constant. In other cases, the system can measure the
G
factor
automatically before an experimental run.
G
factors are incorporated into the
Anisot-
ropy
scan-type:
1
In the FluorEssence main menu, click the Ex-
periment Menu button.
The
Fluorescence Main
Experiment Menu
appears.
2
Click Anisotropy.
The
Experiment Type
window appears.
3
Choose the type of
Anisotropy experiment.
The
Experiment Setup
window opens.
To determine polarization at particular
excitation/emission wavelength-pairs,
choose
vs SinglePoint
.