![horiba FluoroMax-3 Скачать руководство пользователя страница 52](http://html1.mh-extra.com/html/horiba/fluoromax-3/fluoromax-3_operation-manual_2162171052.webp)
FluoroMax-3 v. 3.1 (3 Mar 2006)
System Operation
3-16
Notes on excitation and emission calibration
•
Two experiments,
lamp.exp
and
water.exp
, have been defined and saved.
They can be run, after the system is switched on each day, to check the calibration
and performance of the FluoroMax
®
-3.
•
HORIBA Jobin Yvon Inc. recommends monitoring the number of hours of xenon-
lamp use, via the hour meter.
•
Additionally, you may want to record the water-Raman intensity daily or weekly.
•
The lamp is rated for 1200–1500 h, but if the Raman intensity starts to drop, you
may wish to change the lamp sooner.
Содержание FluoroMax-3
Страница 5: ...FluoroMax 3 v 3 1 3 Mar 2006 v 15 Declaration of Conformity 15 1 16 Index 16 1...
Страница 6: ...FluoroMax 3 v 3 1 3 Mar 2006 vi...
Страница 18: ...FluoroMax 3 v 3 1 3 Mar 2006 Introduction 0 12...
Страница 36: ...FluoroMax 3 v 3 22 Feb 2005 System Description 2 8...
Страница 96: ...FluoroMax 3 v 3 1 3 Mar 2006 Maintenance 6 12...
Страница 156: ...FluoroMax 3 v 3 1 6 Mar 2006 Automated Polarizers 10 22...
Страница 208: ...FluoroMax 3 v 3 1 6 Mar 2006 Bibliography 14 4...
Страница 210: ...FluoroMax 3 v 3 1 6 Mar 2006 Declaration of Conformity 15 2...
Страница 220: ...FluoroMax 3 v 3 1 25 Apr 2006 Index 16 10...