Lamp replacement
The xenon flash lamp typically has a half-intensity life of at least 10 million flashes.
Follow the procedure below for replacement and alignment.
Warning:
High voltage exists within the FluoroMax
®
-P.
To avoid fatal shocks, before removing the lamp cover,
unplug the FluoroMax
®
-P’s power cord, and wait at
least 15 min while the internal capacitors discharge.
Never operate the lamp with the cover removed!
Warning:
Xenon lamps are an explosion hazard. Be
sure that the power is off, and all AC (mains) power is
disconnected from the system. Read and follow all the
cautions below:
Caution:
Intense ultraviolet, visible, or infrared light
may be present when the instrument is open, so
wear eye- and skin-protection, such as light-
blocking goggles and light-blocking clothing.
Содержание FluoroMax-3
Страница 5: ...FluoroMax 3 v 3 1 3 Mar 2006 v 15 Declaration of Conformity 15 1 16 Index 16 1...
Страница 6: ...FluoroMax 3 v 3 1 3 Mar 2006 vi...
Страница 18: ...FluoroMax 3 v 3 1 3 Mar 2006 Introduction 0 12...
Страница 36: ...FluoroMax 3 v 3 22 Feb 2005 System Description 2 8...
Страница 96: ...FluoroMax 3 v 3 1 3 Mar 2006 Maintenance 6 12...
Страница 156: ...FluoroMax 3 v 3 1 6 Mar 2006 Automated Polarizers 10 22...
Страница 208: ...FluoroMax 3 v 3 1 6 Mar 2006 Bibliography 14 4...
Страница 210: ...FluoroMax 3 v 3 1 6 Mar 2006 Declaration of Conformity 15 2...
Страница 220: ...FluoroMax 3 v 3 1 25 Apr 2006 Index 16 10...