FluoroMax-3 v. 3.1 (3 Mar 2006)
Troubleshooting
7-1
7: Troubleshooting
The FluoroMax
®
-3 system has been designed to operate reliably and predictably. If
there is a problem, examine the chart below, and try the steps on the following pages.
Problem Possible
Cause
Remedy
Excitation shutter closed.
Using the software, open the shutter.
Slits are not open to the
proper width.
Adjust the slits.
FluoroMax
®
-P is in phos-
phorimeter layout.
Change to a steady-state layout.
Monochromator is miscali-
brated.
Check and recalibrate monochromator.
Light is not reaching
the sample.
Sample turret is not in cor-
rect position.
Using FluorEssence™, set the position and
open the cover to verify the position.
CW lamp is not aligned.
Align the lamp.
Slits are improperly set.
Adjust the slit widths.
Shutter(s) is(are) not com-
pletely open.
Open the shutter(s).
Lamp power supply is set
to the wrong current rating.
Call the Spex
®
Fluorescence Service Depart-
ment.
Lamp is too old.
Replace lamp. (150-W lamp has lamp lifetime
1200–1500 h.)
Signal intensity is low.
Shutter(s) closed.
Open all shutters.
Signal intensity is at
least 10 times lower
than normal.
Polarizer is in the light
path.
In
Real Time Control
, move the polarizer out
of the light path.
Monochromators are set to
wrong wavelength.
Select appropriate wavelength based on exci-
tation and emission of sample.
Detectors are saturated.
Adjust slits. (Signal detector is linear to 2 ×
10
6
cps in photon-counting mode. Reference
detector saturates at 12 µA.)
Optical density effects and
self-absorption.
Sample is too concentrated. Dilute sample by
a factor of 10 or 100 and retry experiment.
No change in signal
intensity.
Detectors are saturated.
Reduce slit settings.
No signal.
Lamp is not on.
Bad lamp: change xenon lamp.
Lamp unstable.
Let lamp warm up 20 min before use.
Light leaks.
Check dark value to determine.
Erratic signal.
Sample has particles that
scatter light irregularly.
Filter sample, or let particles settle before run-
ning scan.
Содержание FluoroMax-3
Страница 5: ...FluoroMax 3 v 3 1 3 Mar 2006 v 15 Declaration of Conformity 15 1 16 Index 16 1...
Страница 6: ...FluoroMax 3 v 3 1 3 Mar 2006 vi...
Страница 18: ...FluoroMax 3 v 3 1 3 Mar 2006 Introduction 0 12...
Страница 36: ...FluoroMax 3 v 3 22 Feb 2005 System Description 2 8...
Страница 96: ...FluoroMax 3 v 3 1 3 Mar 2006 Maintenance 6 12...
Страница 156: ...FluoroMax 3 v 3 1 6 Mar 2006 Automated Polarizers 10 22...
Страница 208: ...FluoroMax 3 v 3 1 6 Mar 2006 Bibliography 14 4...
Страница 210: ...FluoroMax 3 v 3 1 6 Mar 2006 Declaration of Conformity 15 2...
Страница 220: ...FluoroMax 3 v 3 1 25 Apr 2006 Index 16 10...