FluoroMax-3 v. 3.1 (3 Mar 2006)
Maintenance
6-3
3
Remove the instrument cover.
a
Remove the seven Phillips screws (two on each side, and three in the
back) from the cover.
b
Lift the cover vertically off the instrument by grasping opposite corners
of the cover and raising upward.
If the cover sticks, gently work each side upward until the cover slides smoothly
off.
4
Remove the lamp-housing cover at the rear of
the instrument.
a
Remove the two Phillips screws
from the lamp-housing cover.
b
Gently rotate the lamp-housing
cover with its cooling fans at-
tached.
c
Rotate the cover backwards, and
set it behind the instrument so
that electrical connections are
not strained.
The lamp is held in place by spring ten-
sion and the height adjustment on top of
the lamp. The anode and cathode con-
nections are attached to the lamp via
thumbscrews on top and bottom of the
lamp.
5
Prepare the replacement lamp.
a
Place the new xenon lamp (still in its protective cover) on top of the ex-
citation monochromator.
b
Remove the top half of the new lamp’s protective cover.
Keep this top half handy for later.
6
Remove the old xenon lamp.
Содержание FluoroMax-3
Страница 5: ...FluoroMax 3 v 3 1 3 Mar 2006 v 15 Declaration of Conformity 15 1 16 Index 16 1...
Страница 6: ...FluoroMax 3 v 3 1 3 Mar 2006 vi...
Страница 18: ...FluoroMax 3 v 3 1 3 Mar 2006 Introduction 0 12...
Страница 36: ...FluoroMax 3 v 3 22 Feb 2005 System Description 2 8...
Страница 96: ...FluoroMax 3 v 3 1 3 Mar 2006 Maintenance 6 12...
Страница 156: ...FluoroMax 3 v 3 1 6 Mar 2006 Automated Polarizers 10 22...
Страница 208: ...FluoroMax 3 v 3 1 6 Mar 2006 Bibliography 14 4...
Страница 210: ...FluoroMax 3 v 3 1 6 Mar 2006 Declaration of Conformity 15 2...
Страница 220: ...FluoroMax 3 v 3 1 25 Apr 2006 Index 16 10...