FluoroMax-3 v. 3.1 (3 Mar 2006)
Maintenance
6-5
Warning:
Do not touch any portion of the lamp
except the metal cathode and anode.
posed.
f
Remove the anode cable.
g
Remove the thumbscrew at the cathode end (bottom) of the lamp, leav-
ing the post exposed.
h
Remove the cathode cable.
7
Place the old lamp in the top protective cover
from the new lamp.
8
Put the old lamp (in the top cover) in a safe
place.
9
Attach connections to the new lamp.
a
Attach the cathode connection to the new lamp.
b
Secure the connection with the new lamp’s thumbscrew.
c
Attach the anode connection to the new lamp.
d
Secure the connection with the new lamp’s thumbscrew.
e
Recheck that the connections are correct.
10
Insert the new xenon lamp.
a
Insert the new lamp in the bottom
holder.
b
Press the lamp down into the holder to
compress the spring.
c
Tilt and raise the anode end of the
lamp into the top holder.
d
Place the lamp so that the nipple on
the glass envelope is opposite the
collection mirror.
e
Set the height adjustment, using the
1/8" Allen key. Try to return the new
Anode (+)
Cathode (–)
Col-
lection
mirror
Содержание FluoroMax-3
Страница 5: ...FluoroMax 3 v 3 1 3 Mar 2006 v 15 Declaration of Conformity 15 1 16 Index 16 1...
Страница 6: ...FluoroMax 3 v 3 1 3 Mar 2006 vi...
Страница 18: ...FluoroMax 3 v 3 1 3 Mar 2006 Introduction 0 12...
Страница 36: ...FluoroMax 3 v 3 22 Feb 2005 System Description 2 8...
Страница 96: ...FluoroMax 3 v 3 1 3 Mar 2006 Maintenance 6 12...
Страница 156: ...FluoroMax 3 v 3 1 6 Mar 2006 Automated Polarizers 10 22...
Страница 208: ...FluoroMax 3 v 3 1 6 Mar 2006 Bibliography 14 4...
Страница 210: ...FluoroMax 3 v 3 1 6 Mar 2006 Declaration of Conformity 15 2...
Страница 220: ...FluoroMax 3 v 3 1 25 Apr 2006 Index 16 10...