To insert or remove a sample platform,
1
If a multiple-sample turret is installed, shut off
the system.
2
Remove the four screws on the front of the
sample platform.
3
Slide out the old platform.
4
Slide in the new platform.
5
If the platform has a rotatable turret or magnetic
stirrer, slide the 15-pin connector gently and se-
curely onto the 15-pin receptacle in the sample
compartment.
6
Re-attach the four screws on the front of the
sample platform.
Detectors (5 and 6)
Each FluoroMax
®
-3 contains two detectors:
•
Signal detector
The signal detector is a photon-
counting detector. This detector is
an R928P photomultiplier tube,
which sends the signal to a photon-
counting module. The detector’s
response ranges from 180–850 nm,
with dark counts < 1000 counts per
second (cps). The linear range for photon counting is 0–2 mil-
lion cps. The working range is up to about 1–2 million cps.
Above 2 million cps, photon-pulse pile-up reduces the signal-
to-noise ratio, and causes the detector to lose its linear re-
sponse.
•
Reference detector
The reference detector monitors the
xenon lamp, in order to correct for
wavelength- and time-dependent
output of the lamp. This detector is a
UV-enhanced silicon photodiode,
which is just before the sample
compartment. It requires no external bias, and has good re-
sponse from 190–980 nm. The output is collected by a cur-
rent-input module, with a range from 0–12 µA.
Содержание FluoroMax-3
Страница 5: ...FluoroMax 3 v 3 1 3 Mar 2006 v 15 Declaration of Conformity 15 1 16 Index 16 1...
Страница 6: ...FluoroMax 3 v 3 1 3 Mar 2006 vi...
Страница 18: ...FluoroMax 3 v 3 1 3 Mar 2006 Introduction 0 12...
Страница 36: ...FluoroMax 3 v 3 22 Feb 2005 System Description 2 8...
Страница 96: ...FluoroMax 3 v 3 1 3 Mar 2006 Maintenance 6 12...
Страница 156: ...FluoroMax 3 v 3 1 6 Mar 2006 Automated Polarizers 10 22...
Страница 208: ...FluoroMax 3 v 3 1 6 Mar 2006 Bibliography 14 4...
Страница 210: ...FluoroMax 3 v 3 1 6 Mar 2006 Declaration of Conformity 15 2...
Страница 220: ...FluoroMax 3 v 3 1 25 Apr 2006 Index 16 10...