FluoroMax-3 v. 3.1 (3 Mar 2006)
Troubleshooting
7-6
The following lamp scan spectrum shows poor resolution in the area around the peak.
Poor lamp scan of 150-W Xe lamp. Note low resolution in the area near
the 467-nm peak.
This lack of spectral resolution appears because the slit widths are set too wide. To re-
solve this problem, narrow the slit widths.
Intensity (cou
nts
s
–
1
)
0.4473
0.3355
0.2236
0.1118
0
Xenon-lamp peaks
are unresolved
300
400
500 600 700 800
Wavelength (nm)