FluoroMax-3 v. 3.1 (3 Mar 2006)
System Operation
3-10
Note:
Avoid glass or acrylic cu-
vettes: they may exhibit UV fluo-
rescence or filtering effects.
Emission calibration check
This calibration check verifies the wavelength calibration of the emission monochroma-
tor with the emission photomultiplier tube. It is an emission scan of the Raman-scatter
band of water performed in right-angle mode. Perform this check after the xenon-lamp
scan. When completed, the performance of the system has been verified.
The water sample should be research-quality, triple-distilled or deionized water. HPLC-
grade (18-M
Ω
spec.) or equivalent water is suggested for the Raman scan. Impure sam-
ples of water will cause elevated background levels as well as distorted spectra with
(perhaps) some unwelcome peaks.
Use a 4-mL quartz cuvette.
1
Insert the water
sample into the
sample compartment.
With an automated sample changer, note the position number in which the sam-
ple cell is placed.
2
Close the lid of the sample chamber.
3
In the main
FluorEssence
toolbar, choose the Ex-
periment Menu button.
The
Fluorescence Main Experiment Window
opens.
Note:
The emission calibration of the instrument is directly affected by
the calibration of the excitation monochromator.
Содержание FluoroMax-3
Страница 5: ...FluoroMax 3 v 3 1 3 Mar 2006 v 15 Declaration of Conformity 15 1 16 Index 16 1...
Страница 6: ...FluoroMax 3 v 3 1 3 Mar 2006 vi...
Страница 18: ...FluoroMax 3 v 3 1 3 Mar 2006 Introduction 0 12...
Страница 36: ...FluoroMax 3 v 3 22 Feb 2005 System Description 2 8...
Страница 96: ...FluoroMax 3 v 3 1 3 Mar 2006 Maintenance 6 12...
Страница 156: ...FluoroMax 3 v 3 1 6 Mar 2006 Automated Polarizers 10 22...
Страница 208: ...FluoroMax 3 v 3 1 6 Mar 2006 Bibliography 14 4...
Страница 210: ...FluoroMax 3 v 3 1 6 Mar 2006 Declaration of Conformity 15 2...
Страница 220: ...FluoroMax 3 v 3 1 25 Apr 2006 Index 16 10...