FluoroMax-3 v. 3.1 (3 Mar 2006)
Maintenance
6-2
Hazards
!
Xenon-arc lamps are an explosion
hazard. Wear explosion-proof face-
shield and protective clothing when
opening the lamp housing and han-
dling the lamp.
!
Disconnect the lamp power supply from the AC power line (mains) while handling
lamp leads. Lethal high voltages may exist.
!
The lamp remains extremely hot for approximately one-half hour after it has been
turned off. Do not touch the lamp or the metal unit until the lamp has cooled.
!
Never look directly at the xenon arc or its reflection. Intense radiation can perma-
nently damage eyes.
!
Do not touch the focusing lens, back-scatter mirror, or the surface of the lamp. Fin-
gerprints will be burned onto these surfaces when the lamp is ignited.
Changing the lamp
1
Switch off and prepare the FluoroMax
®
-3.
a
Be sure that the FluoroMax
®
-3 and the host computer are turned off.
b
Remove the AC (mains) power cord from the FluoroMax
®
-3.
c
Disconnect the RS-232 cable, optional trigger-box cable, power cord,
and any other cables attached to the spectrofluorometer.
2
Remove the sample mount from the front of the
FluoroMax
®
-3.
a
Remove the four Phillips screws that secure the sample mount to the in-
strument.
b
Gently slide the sample mount out of the instrument.
Some sample mounts have a 15-pin connector at the inside end for automated
accessories.
Содержание FluoroMax-3
Страница 5: ...FluoroMax 3 v 3 1 3 Mar 2006 v 15 Declaration of Conformity 15 1 16 Index 16 1...
Страница 6: ...FluoroMax 3 v 3 1 3 Mar 2006 vi...
Страница 18: ...FluoroMax 3 v 3 1 3 Mar 2006 Introduction 0 12...
Страница 36: ...FluoroMax 3 v 3 22 Feb 2005 System Description 2 8...
Страница 96: ...FluoroMax 3 v 3 1 3 Mar 2006 Maintenance 6 12...
Страница 156: ...FluoroMax 3 v 3 1 6 Mar 2006 Automated Polarizers 10 22...
Страница 208: ...FluoroMax 3 v 3 1 6 Mar 2006 Bibliography 14 4...
Страница 210: ...FluoroMax 3 v 3 1 6 Mar 2006 Declaration of Conformity 15 2...
Страница 220: ...FluoroMax 3 v 3 1 25 Apr 2006 Index 16 10...