![horiba FluoroMax-3 Скачать руководство пользователя страница 153](http://html1.mh-extra.com/html/horiba/fluoromax-3/fluoromax-3_operation-manual_2162171153.webp)
FluoroMax-3 v. 3.1 (6 Mar 2006)
Automated Polarizers
10-19
Warning:
Refer to the Materials Safety Data Sheet
(MSDS) for detailed information on methanol.
Maintenance
Like all optics, polarizers should be handled with care and stored properly. With proper
care, a polarizer should last for many years. Aside from installation, removal, and stor-
age, there is no routine maintenance necessary for a polarizer. Polarizers should be re-
moved and stored when not in use. Store the polarizers in their collars to maintain cali-
bration, in a drawer or cabinet. Wrap the polarizers in lens tissue—to keep them dust-
free and for protection—and then place them in a plastic bag. The automated accesso-
ries should also be stored in a dust-free environment.
Should the polarizer windows need cleaning, apply a mild solution of methanol, and
blow it dry.
We recommend measuring the anisotropy of scatter (to verify the alignment of the crys-
tals) before any critical experiment. In addition to the standard xenon-lamp spectrum
and water Raman spectra, which serve to verify the wavelength calibration, measure-
ment of the anisotropy of scatter will provide a fast check that the instrument system is
ready to perform measurements.
Содержание FluoroMax-3
Страница 5: ...FluoroMax 3 v 3 1 3 Mar 2006 v 15 Declaration of Conformity 15 1 16 Index 16 1...
Страница 6: ...FluoroMax 3 v 3 1 3 Mar 2006 vi...
Страница 18: ...FluoroMax 3 v 3 1 3 Mar 2006 Introduction 0 12...
Страница 36: ...FluoroMax 3 v 3 22 Feb 2005 System Description 2 8...
Страница 96: ...FluoroMax 3 v 3 1 3 Mar 2006 Maintenance 6 12...
Страница 156: ...FluoroMax 3 v 3 1 6 Mar 2006 Automated Polarizers 10 22...
Страница 208: ...FluoroMax 3 v 3 1 6 Mar 2006 Bibliography 14 4...
Страница 210: ...FluoroMax 3 v 3 1 6 Mar 2006 Declaration of Conformity 15 2...
Страница 220: ...FluoroMax 3 v 3 1 25 Apr 2006 Index 16 10...