FluoroMax-3 v. 3.1 (7 Mar 2006)
Components & Accessories
12-31
Caution:
Never touch any optical surfaces of the
gratings, mirrors, etc.
Note:
The window
assemblies will be
mounted here.
Caution:
Take care to avoid contact with the
beam-splitter (with one window assembly)
and the optics (with the other assembly).
Note:
If there are no filter holders,
mount the FM-2007 Windows on
the inside of the sample compart-
ment. Go directly to step 2a.
FM-2007 Windows for the FluoroMax
®
-3
sample compartment
Installing the windows
1
Prepare the
instrument.
a
Remove the instrument cover.
b
Remove the two black irregularly-shaped covers over the optical path.
2
Insert window
assemblies.
a
Snap one window
assembly onto one
external side-wall of
the sample compartment.
b
Insert 6-32 × 3/8" cap screw.
c
Tighten the cap screw.
d
Repeat steps (a) through (c) for the other window assembly.
Beam-
splitter