FluoroMax-3 v. 3.1 (7 Mar 2006)
Components & Accessories
12-27
SFA-20/SPEX Stopped-flow accessory
The SFA-20/SPEX series of
stopped-flow rapid-kinetics
accessories offers versatility
for spectroscopic monitoring of
fast reactions in solution.
In addition to the conventional
two-syringe mixing system,
there is also a three-syringe
version with two sequential
mixers in the cell, giving you
an option to do double-mixing.
Further choices include a
micro-volume version,
reducing the volumes of
reagents required to load the
instrument, and thus improve
sample economy.
The SFA-series stopped-flow accessory permits observation of the reaction rate of two
reactants forced through a mixing chamber, and into an observation cell. The reactant
solutions are contained in drive syringes whose pistons simultaneously are driven. After
leaving the observation cell, the reactants advance a stop syringe, triggering data-
acquisition by the spectrofluorometer.
This instrument has been designed to suit the particular needs of FluoroMax
®
spectro-
fluorometers. The optical cell matches the beam geometry of this instrument. A cable is
supplied so data-acquisition can be externally triggered at stopping, providing a repro-
ducible time-zero registration for all traces, and allowing accurate overlay and averag-
ing.
Содержание FluoroMax-3
Страница 5: ...FluoroMax 3 v 3 1 3 Mar 2006 v 15 Declaration of Conformity 15 1 16 Index 16 1...
Страница 6: ...FluoroMax 3 v 3 1 3 Mar 2006 vi...
Страница 18: ...FluoroMax 3 v 3 1 3 Mar 2006 Introduction 0 12...
Страница 36: ...FluoroMax 3 v 3 22 Feb 2005 System Description 2 8...
Страница 96: ...FluoroMax 3 v 3 1 3 Mar 2006 Maintenance 6 12...
Страница 156: ...FluoroMax 3 v 3 1 6 Mar 2006 Automated Polarizers 10 22...
Страница 208: ...FluoroMax 3 v 3 1 6 Mar 2006 Bibliography 14 4...
Страница 210: ...FluoroMax 3 v 3 1 6 Mar 2006 Declaration of Conformity 15 2...
Страница 220: ...FluoroMax 3 v 3 1 25 Apr 2006 Index 16 10...