FluoroMax-3 v. 3.1 (3 Mar 2006)
System Operation
3-6
Note:
Your lamp scan may appear different,
depending on the instrument configuration.
This is an
uncalibrated
FluoroMax
®
lamp scan. The
main peak
ought to be at 467 nm, but here appears near 480 nm.
7
Calibrate the excitation monochromator, if re-
quired.
a
Expand the plot by clicking the
Expand
button.
b
Click the cursor button to start the
Cursor
function.
c
Click on the graph near the peak, to place the cursor on the graph.
d
Using the left and right arrows on the keyboard, move the cursor to the
top of the peak.
e
Read the
x
-value of this plot: this is the wavelength of the peak.
structure ~ 450 nm
calibration
peak at 467
broadband