FluoroMax-3 v. 3.1 (3 Mar 2006)
Optimizing Data
5-16
c
Choose
Preferences
, then the
Instrument Correction Files
icon.
The
Instrument Correction Files
area should display a correction file for the
Detector
(
S
or
R
). If not, click the
Insert
button, and browse for the desired
correction file.
Detector
Correction-factor file name
S
mcorrect
R
xcorrect
T
(if used)
tcorrect
2
Choose correction when setting up the experi-
ment.
a
In the
Experiment Setup
window, choose the
Detectors
icon
to display the detectors’ parameters.
b
Click the
Correction
checkbox for the detector you want corrected.
In the
Signal Algebra
area, a signal with appended “
c
” appears, denoting a cor-
rected signal: