FluoroMax-3 v. 3.1 (3 Mar 2006)
Troubleshooting
7-10
Further assistance...
Read all software and accessory manuals before contacting the Spex
®
Fluorescence
Service Department. Often the manuals show the problem’s cause and a method of so-
lution. Technical support is available for both hardware and software troubleshooting.
Before contacting the service department, however, complete the following steps.
1
If this is the first time the problem has occurred,
try turning off the system and accessories.
After a cool-down period, turn everything back on.
2
Make sure all accessories are properly
configured, and turned on as needed.
3
Following the instructions in Chapter 3,
System
Operation
, run a lamp scan and a water Raman
scan to make sure the system is properly
calibrated.
Print the spectrum for each and note the peak intensities.
4
Check this chapter to see if the problem is
discussed.
5
Try to duplicate the problem and write down the
steps required to do so.
The service engineers will try to do the same with a test system. Depending on
the the problem, a service visit may not be required.
6
If an error dialog box appears in
FluorEssence™, write down the exact error
displayed.
7
In FluorEssence™, in the
FluorEssence
main
window’s
toolbar, choose Help.
A drop-down menu appears.
Содержание FluoroMax-3
Страница 5: ...FluoroMax 3 v 3 1 3 Mar 2006 v 15 Declaration of Conformity 15 1 16 Index 16 1...
Страница 6: ...FluoroMax 3 v 3 1 3 Mar 2006 vi...
Страница 18: ...FluoroMax 3 v 3 1 3 Mar 2006 Introduction 0 12...
Страница 36: ...FluoroMax 3 v 3 22 Feb 2005 System Description 2 8...
Страница 96: ...FluoroMax 3 v 3 1 3 Mar 2006 Maintenance 6 12...
Страница 156: ...FluoroMax 3 v 3 1 6 Mar 2006 Automated Polarizers 10 22...
Страница 208: ...FluoroMax 3 v 3 1 6 Mar 2006 Bibliography 14 4...
Страница 210: ...FluoroMax 3 v 3 1 6 Mar 2006 Declaration of Conformity 15 2...
Страница 220: ...FluoroMax 3 v 3 1 25 Apr 2006 Index 16 10...