FluoroMax-3 v. 3.1 (3 Mar 2006)
System Operation
3-9
i
Enter the current, observed position of the peak in the
Position Control
(here, 477 nm).
j
Click the
Calibrate Excitation 1
button.
The
Calibrate
window
opens:
k
In
Peak Of Interest
, enter the actual or expected position of the peak (it
ought to be 467 nm).
l
Click
OK
.
m
At the bottom right of the
Real Time Control
window, click
Cancel
.
n
In the
Experiment
window, click
Run
to confirm the correct peak posi-
tion. A correct scan is shown below (peak is at 467 nm).
Cursor
Wave-
length in
nm