3.2.4
3 - 87
(cont’d)
Setting Item
Description
Sensitivity
Set up a condition for detecting a peak and valley
from the measured spectrum.
It is sensitivity that determines the detectability in the
X-axis direction of a spectrum. For detection of a
sharp peak, select sensitivity 1. For detection of a
broad peak, select sensitivity 8. The following 4
levels are selectable.
[1] 1
[2] 2
[3] 4
[4] 8
For details, refer to Explanation 3-9.
(3) After completion of calculation condition setting, return
to the WL Scan screen by pressing the [RETURN] or
[
◄
] key.