3.3 Sample-by-sample Measurement (6-cell manual mode)
9. Baseline
Correction
(1) To determine the measuring conditions set in the
preceding steps, press the <8> key (Meas. Screen)
and then [ENTER] key. Or press the [MEAS.
SCREEN] key.
(2) The guidance on baseline correction (Fig. 3-86)
appears. Set a sample for baseline correction to cell
A. Upon pressing the [START] key, baseline
correction is carried out. If there is a baseline saved
under the same measuring conditions, you can also
proceed to sample measurement without performing
baseline correction with the CLEAR key.
WL Scan
CELL A
Baseline CORR.
[Info]
Set baseline sample to Cell A.
START:Baseline CORR.
Fig. 3-86 Guidance on Baseline Correction
10. Measurement of Sample Solutions
(1) After completion of baseline correction, the sample
measurement screen (Fig. 3-87) appears. Set a
sample at the current cell position and press the
[START] key. Measurement will then start.
For performing baseline correction, press the
[RETURN] key.
3 - 130