5.2 Check with Pen Type Low-Pressure Mercury Lamp Available at Option
5.2.4 Report Printing
(1) Display the Performance Check (Option) screen (Fig. 5-42).
For display method, follow the procedure starting from
section 5.2.. By the Print Report function, each item
subjected to performance check with the Hg lamp can be
printed in the format of a report. For execution of this
function, select [1] Yes.
Performance
CELL A
(Option)
WL Accuracy Hg (253.7 nm)
WL Accuracy Hg (435.8 nm)
WL Accuracy Hg (546.1 nm)
WL Repeatability Hg
Resolution Hg (546.1nm
Hardware
Print Report
Execute it? :_
□
1 Yes □
2 No
Fig. 5-42 Performance Check (Option) Screen
(2) A report is printed as shown below.
Report: 2010/04/10 15:12
Model
: U-5100 Spectrophotometer
Serial No.
: 1234-567
ROM VER.
: 3J25300
Option
:
Performance Check
WL Accuracy Hg (253.7 nm): -0.4 nm Tolerance
±
2.0 nm OK
Items subjected to
check are printed in
this area.
option performance
5 - 32