3.3 Sample-by-sample Measurement (6-cell manual mode)
(6) When the specified number of standards have all been
measured, the screen shown in Fig. 3-68 appears.
When the <1> key (Sample Meas.) is pressed, the
system terminates standard measurement and
proceeds to sample measurement. Pressing the <2>
key (Curve) or <3> key (Save) allows the processing
shown in Table 3-44.
Photometry
CELL 1
ID
ABS
CONC
STD1
0.001
0.000
STD2
0.102
10.00
STD3
0.205
20.00
Select function. : _
□
1 Sample Meas. □
2 Curve □
3 Save □
4 Print
Fig. 3-68 Screen after Measurement of All Standards
3 - 106