3.3 Sample-by-sample Measurement (6-cell manual mode)
(cont’d)
Setting Item
Description
[4] Print
You can select items to be printed.
When this function is selected, the screen
given below appears. For each item,
specify whether to print out (ON or OFF).
Photometry
CELL A
Print Date
ON
Run Date
ON
Method
ON
Curve
ON
STD
ON
Sample
ON
Select function. : _
□
1 ON □
2 OFF
10. Measurement of Sample Solutions
(1) The sample measurement screen (Fig. 3-69) appears.
Set the sample indicated in the guidance area in the
sample compartment. Pressing the [START] key
starts measurement at the current cell position.
NOTE
:
In 6-cell manual mode, the photometric value at
the cell position indicated on the screen is
measured. Before measurement, therefore,
check if the cell position where the sample has
been set coincides with the cell position
indicated on the screen.
Photometry
CELL 1
ID
ABS
CONC
1
START:Meas.SAMP.1; Select FUNC..:_
□
1 Curve Confirmation
Fig. 3-69 Sample Measurement Screen