3.3 Sample-by-sample Measurement (6-cell manual mode)
3.3.4 Measuring a Spectrum
This instrument is capable of measuring the transmission spectrum and absorption spectrum.
By setting the blank to cell A, one sample can be measured after automatic execution of
baseline correction.
GUIDE
: When auto start is specified, the system automatically sets up the conditions and
advances to the measurement screen phase of sequence after the power switch is
turned on. For how to specify auto start, refer to section 4.5..
1.
Startup of This Product
Start this product. (For the starting procedure, refer to
section 2.3.1..)
*
2.
Setting of Measuring Conditions
3.
Setting of Sample Conditions
* For steps 2 and 3, refer to section 3.2.4..
4.
Setting of 6-cell Conditions
NOTE
:
“6 Cell Mode” is not indicated if OFF is specified
for Mode on the 6-cell mode setting screen for
use of the single cell holder or rectangular long
path cell holder. (For details, refer to section
2.5.4..) When the auto sipper is connected,
“Sipper” is indicated in place of “6 Cell Mode”.
(1) To set the 6-cell conditions, press the <3> key (6 Cell)
and then [ENTER] key. Or select “6 Cell” with the [
▲
]
or [
▼
] key and press the [ENTER] key or [
►
] key.
3 - 128
Содержание U-5100
Страница 280: ...5 1 Check with Instrument Main Unit Alone Fig 5 6 Example Result of Wavelength Accuracy Check at 484 6 nm 5 6 ...
Страница 283: ...5 1 2 Fig 5 10 Example Result of Wavelength Repeatability Check 5 9 ...
Страница 286: ...5 1 Check with Instrument Main Unit Alone Fig 5 14 Example Result of Noise Level Check 5 12 ...
Страница 289: ...5 1 4 Fig 5 18 Example Result of Baseline Flatness Check 5 15 ...
Страница 292: ...5 1 Check with Instrument Main Unit Alone Fig 5 22 Example Result of Baseline Stability Check 5 18 ...