3.2.2
3 - 55
Table 3-25 Auto Zero Interval and Operation of Instrument
SAMP. Autozero: ON
SAMP. Autozero: OFF Auto Zero
Interval: 5
Auto Zero
Interval: 1
Cell Position
Operation
Setting of samples (1st round)
⎯
1
1
Cell A
Auto zero
1
2
2
Cell 1
Sample 1 measurement
⎯
⎯
3
Cell A
Auto zero
2
3
4
Cell 2
Sample 2 measurement
⎯
⎯
5
Cell A
Auto zero
3
4
6
Cell 3
Sample 3 measurement
⎯
⎯
7
Cell A
Auto zero
4
5
8
Cell 4
Sample 4 measurement
⎯
⎯
9
Cell A
Auto zero
5
6
10
Cell 5
Sample 5 measurement
Setting of samples (2nd round)
⎯
7
11
Cell A
Auto zero
6
8
12
Cell 1
Sample 6 measurement
⎯
⎯
13
Cell A
Auto zero
7
9
14
Cell 2
Sample 7 measurement
Repeated until the specified number of samples are all measured.
When OFF is set for “SAMP. Autozero”, auto zero can
be executed by pressing the [AUTOZERO] key when
the guidance on sample setting is indicated.
For details, refer to Fig. 3-31.