6.6 Replacement of Power Fuse
WARNING
Before replacement, make sure that the power cord is not
connected.
If the power fuse is blown due to any cause, it should be
replaced with a new one in the following procedure. If the new
fuse is also blown, the instrument may be faulty. Contact your
dealer or the maintenance service company authorized by
Hitachi High-Technologies.
(1) Unplug the power cord from the connector of the
spectrophotometer. (See <1> in Fig. 6-4.)
<1>
Fig. 6-4 Unplugging the Power Cord
(2) The fuse holder is located above the power connector on
the rear side of spectrophotometer. Push the fuse holder
stopper from either side using a blade-edge screwdriver,
and the holder will come out slightly. (See <2> in Fig.
6-5.) In succession, push the other stopper in the same
manner as above, and the entire holder will advance slightly
toward you. Then, grasp and pull out the holder.
6 - 5