6.8 If You Encounter a Trouble
(cont’d)
Error Message
Cause
Remedy
WL Motor Error
Irregularity is found
when driving the
wavelength motor.
Retry driving. If no
improvement can be
seen, contact the
maintenance service
company authorized by
Hitachi High-
Technologies.
Capacitor Error
Irregularity is found
when setting the
capacitor.
Retry setting. If no
improvement can be
seen, contact the
maintenance service
company authorized by
Hitachi High-
Technologies.
b. Troubleshooting
Symptom Cause
Remedy
No indication
appears even if the
power switch is
turned on.
<1> The power cord
is unplugged.
<2> The fuse is
blown.
<3> Any cause other
than above
<1> Plug in the power
cord.
<2> Replace the fuse
with a new one.
<3> If the same
symptom persists
after tuning off the
power switch and
turning it on, contact
the maintenance
service company
authorized by
Hitachi High-
Technologies.
Significant
variations in
measured values
<1> The sample cell
or the
transparent
plate is
contaminated or
clouded with
water drops.
<2> The lid of
sample
compartment is
not closed.
<3> Any cause other
than above
<1> Remove
contaminants or
water drops.
<2> Close the lid of
sample
compartment.
<3> If the same
symptom persists
after tuning off the
power switch and
turning it on, contact
the maintenance
service company
authorized by
Hitachi High-
Technologies.
6 - 12
Содержание U-5100
Страница 280: ...5 1 Check with Instrument Main Unit Alone Fig 5 6 Example Result of Wavelength Accuracy Check at 484 6 nm 5 6 ...
Страница 283: ...5 1 2 Fig 5 10 Example Result of Wavelength Repeatability Check 5 9 ...
Страница 286: ...5 1 Check with Instrument Main Unit Alone Fig 5 14 Example Result of Noise Level Check 5 12 ...
Страница 289: ...5 1 4 Fig 5 18 Example Result of Baseline Flatness Check 5 15 ...
Страница 292: ...5 1 Check with Instrument Main Unit Alone Fig 5 22 Example Result of Baseline Stability Check 5 18 ...