3.2 Automatic Continuous Measurement (6-cell auto mode)
3 - 50
NOTE
:
•
If a file of the same name as entered already exists in
the same measurement mode, the following guidance
appears.
Already exists. Overwrite it? :_
□
1 Yes □
2 No
For overwriting, press the <1> key (Yes). To avoid
overwriting, press the <2> key (No) and rename the file
to be saved.
•
A total of 50 method files can be saved for all
measurement modes.
•
If 50 files are already registered, the following message
appears.
No.of files is full.
In this case, delete files referring to section 4.2.2 and
retry saving.
•
The method for which auto start (for details, refer to
section 4.5.) is set is preceded by “*”.
(5) Return to the Photometry screen by pressing the
[RETURN] key.