3.3.1
Table 3-45 Guidance after Measurement of All
Samples
Setting Item
Description
[1] Meas.
You can make setting related to
measurement. “New Meas.”, “Continue”
and “Remeasure” are selectable.
[1] New Meas. : Selected when newly
measuring samples
with the same
calibration curve.
[2] Continue
: Allows you to start
measurement from the
number following the
previously measured
sample. When this
function is selected, the
guidance given below
appears. Enter the
number of samples to
be measured and press
the [ENTER] key.
Input value. : _
No.of Sample [1-140]
Display With 10 Samples Measured
[3] Remeasure : Selected when
remeasuring a
measured sample.
When this function is
selected, the guidance
given below appears.
Select a sample to be
remeasured with the
[
▲
] or [
▼
] key, set the
sample to cell 1 and
press the [START] key
to carry out remeasure-
ment. To execute
auto zero, press the
[AUTOZERO] key.
Photometry
CELL A
ID
ABS
CONC
6
0.301
30.00
7
0.251
25.11
8
0.419
40.90
9
0.110
11.00
10
0.234
23.50
Set sample to Cell1(Sample10).: _
START:Meas.; AUTOZERO:Autozero
◄
►
( 2/ 2)
3 - 111
Содержание U-5100
Страница 280: ...5 1 Check with Instrument Main Unit Alone Fig 5 6 Example Result of Wavelength Accuracy Check at 484 6 nm 5 6 ...
Страница 283: ...5 1 2 Fig 5 10 Example Result of Wavelength Repeatability Check 5 9 ...
Страница 286: ...5 1 Check with Instrument Main Unit Alone Fig 5 14 Example Result of Noise Level Check 5 12 ...
Страница 289: ...5 1 4 Fig 5 18 Example Result of Baseline Flatness Check 5 15 ...
Страница 292: ...5 1 Check with Instrument Main Unit Alone Fig 5 22 Example Result of Baseline Stability Check 5 18 ...