4.3 Data Processing
(2) Each standard data is now displayed as shown below.
Photometry/Curve
CELL A
ID
ABS
CONC
STD1
0.001
0.000
STD2
0.099
10.00
STD3
0.203
20.00
STD4
0.300
30.00
STD5
0.405
40.00
Select function. :_
[1] Remeasure [2] Delete [3] Recover
Fig. 4-34 Each Standard Data
Table 4-14 Guidance for Photometry/Curve
Function Key
Explanation
[1] Remeasure
To be selected for remeasuring an already
measured sample.
Upon selection, the following guidance
appears. Select a sample to be
remeasured by the [
▲
]/[
▼
] key and place
the sample in cell 1. Then, press the [Start]
key. Remeasurement will start. For auto
zero operation, press the [AUTOZERO] key.
Photometry
CELL A
ID
ABS
CONC
6
0.301
30.00
7
0.251
25.11
8
0.419
40.90
9
0.110
11.00
10
0.234
23.50
Set sample to Cell1(Sample 10).: _
START:Meas.; AUTOZERO:Autozero
◄
►
( 2/ 2)
[2] Delete
The measured data of a standard which you
don’t want to use for a calibration curve can
be excluded from calculation by selecting the
data by [
▲
]/[
▼
] key and pressing the <1>
key (for Delete).
After deletion, the asterisk (*) mark is put at
the head of the data.
Select function. :_
[1]Delete
[3] Recover
The measured data of a standard which has
been excluded from calculation can be
recovered by selecting the data by [
▲
]/[
▼
]
key and pressing the <1> key (for Recover).
After recovery, the asterisk (*) mark put at
the head of the data will disappear.
Select function. :_
[1] Recover
4 - 30
Содержание U-5100
Страница 280: ...5 1 Check with Instrument Main Unit Alone Fig 5 6 Example Result of Wavelength Accuracy Check at 484 6 nm 5 6 ...
Страница 283: ...5 1 2 Fig 5 10 Example Result of Wavelength Repeatability Check 5 9 ...
Страница 286: ...5 1 Check with Instrument Main Unit Alone Fig 5 14 Example Result of Noise Level Check 5 12 ...
Страница 289: ...5 1 4 Fig 5 18 Example Result of Baseline Flatness Check 5 15 ...
Страница 292: ...5 1 Check with Instrument Main Unit Alone Fig 5 22 Example Result of Baseline Stability Check 5 18 ...