2.5.5
(2) After completion of setting, return to the Main Menu screen
with the [RETURN] key.
Note
:
When Method is opened from Method Menu (Refer to
section 4.2.1 for details), Auto Start (Refer to section 4.5 for
details), and data is opened from Data Menu (Refer to section
4.1.1 for details), the setting of ON/OFF of A is subjected to
change.
2.5.5 Selection of Language
You can select a language on the data display between English
and Japanese in the following procedure.
1.
Startup of Instrument
Start the instrument according to section 2.3.1.
2.
Main Menu Screen
The Main Menu screen (Fig. 2-32) appears. To select
“System Setup”, press the <6> key (System Setup) and
then [ENTER] key. Or select “System Setup” with the [
▲
]
or [
▼
] key and press the [ENTER] key.
Main Menu
CELL A
1 Photometry
2 WL Scan
3 Time Scan
4 Method Menu
5 Data Menu
6 System Setup
7 Performance
8 Data Display
Select function. :_
Fig. 2-32 Main Menu Screen
3.
System Setup Screen
The System Setup screen (Fig. 2-33) appears. To select
“Language”, press the <7> key (Language) and then
[ENTER] key. Or select “Language” with the [
▲
] or [
▼
]
key and press the [ENTER] key.
2 - 31
Содержание U-5100
Страница 280: ...5 1 Check with Instrument Main Unit Alone Fig 5 6 Example Result of Wavelength Accuracy Check at 484 6 nm 5 6 ...
Страница 283: ...5 1 2 Fig 5 10 Example Result of Wavelength Repeatability Check 5 9 ...
Страница 286: ...5 1 Check with Instrument Main Unit Alone Fig 5 14 Example Result of Noise Level Check 5 12 ...
Страница 289: ...5 1 4 Fig 5 18 Example Result of Baseline Flatness Check 5 15 ...
Страница 292: ...5 1 Check with Instrument Main Unit Alone Fig 5 22 Example Result of Baseline Stability Check 5 18 ...