4.4 Measurement in Statistical Calculation Mode
[For measurement of DNA sample (by ratio calculation)]
Depending on each mode, set measurement parameters by
reference to the following subsections. However, in setting of
sample conditions, you must set the statistical calculation in ON
status, and set the number of samples for statistical calculation
in setting the number of calculations.
6-cell auto mode:
3.2.3 Measuring DNA Sample (measurement through ratio
calculation)
6-cell manual mode, 6-cell mode OFF:
3.3.3 Measuring DNA Sample (measurement through ratio
calculation)
After measurement, the result of statistical calculation is
displayed every specified number of calculations as shown
below. N, MEAN, SD and RSD stand for the number of
calculations, mean absorbance difference or ratio value,
standard deviation and relative standard deviation, respectively.
Photometry 280.0 nm 0.268 ABS
CELL A
ID
ABS
2
WL1
( 260.0 nm)
0.423
WL2
( 280.0 nm)
0.268
N=2
[MEAN] [SD]
[RSD]
ABS DIFF = 0.155
[0.155]
[0.003]
[0.1]
ABS Ratio= 1.582
[1.582]
[0.004]
[0.2]
Fig. 4-61 Result of Statistical Calculation in
DNA Measurement
4 - 48
Содержание U-5100
Страница 280: ...5 1 Check with Instrument Main Unit Alone Fig 5 6 Example Result of Wavelength Accuracy Check at 484 6 nm 5 6 ...
Страница 283: ...5 1 2 Fig 5 10 Example Result of Wavelength Repeatability Check 5 9 ...
Страница 286: ...5 1 Check with Instrument Main Unit Alone Fig 5 14 Example Result of Noise Level Check 5 12 ...
Страница 289: ...5 1 4 Fig 5 18 Example Result of Baseline Flatness Check 5 15 ...
Страница 292: ...5 1 Check with Instrument Main Unit Alone Fig 5 22 Example Result of Baseline Stability Check 5 18 ...