4.6.3
4.6.3 Rectangular Long-Path Cell Holder (Separately Available)
The rectangular long-path cell holder enables the user to
measure a sample using a cell having an optical path length
within 10 to 100 mm. In this application, low-concentration
samples can be measured at enhanced absorbance levels.
1.
Removal of 6-cell Holder
Remove the 6-cell holder, referring to the 6-cell holder
removing method described in “a.” of 2.4.5.
2.
Mounting of Rectangular Long-path Cell Holder
(1) Open the sample compartment and place the
rectangular long-path cell holder so that its positioning
holes receive the guide pins of the sample
compartment. This holder must be mounted locating
the rectangular cell mounting position on the front side
of the instrument as shown below.
Guide pins
Fig. 4-73 Guide Pins of Sample Compartment
Positioning holes
Locate on the front side
of instrument.
Fig. 4-74 Appearance of Rectangular Long-Path
Cell Holder
4 - 57