3.3.4
(2) The 6-cell condition (6 Cell) screen (Fig. 3-85)
appears. To set each item on the 6-cell condition (6
Cell) screen, press the [ENTER] key. Or use the [
▲
]
or [
▼
] key. Select “Manual” for 6 Cell Mode and
specify the number of samples to be measured for No.
of Sample. For details of each parameter, refer to
Table 3-50.
WL Scan
CELL A
1 Meas.PARAM.
6 Cell Mode
Auto
2 Sample
No.of Sample
1
3 6 Cell
4 System
5 Peak PARAM.
6 Print
7 Save Method
8 Meas.Screen
Input character. :
Sample Name [Max: 8 characters]
Fig. 3-85 6-cell Condition (6 Cell) Screen
Table 3-50 “6 Cell” Setting Parameter
Setting Item
Description
6 Cell Mode
Select a 6-cell operation mode for
measurement from the following.
1) Auto
By setting a baseline sample and
unknown sample to the 6-cell holder, it
is possible to automatically turn the 6-
cell holder and automatically measure
one sample after baseline correction.
(For details of the Auto mode, refer to
section 3.2.4..)
2) Manual
Baseline correction and sample
measurement are performed manually.
It is required to manually move the cells
and execute baseline correction.
5.
Setting of System Conditions
6.
Setting of Peak Detection Conditions
*
7.
Setting of Printing Conditions
8.
Saving of Measuring Conditions
* For steps 5 to 8, refer to section 3.2.4..
3 - 129
Содержание U-5100
Страница 280: ...5 1 Check with Instrument Main Unit Alone Fig 5 6 Example Result of Wavelength Accuracy Check at 484 6 nm 5 6 ...
Страница 283: ...5 1 2 Fig 5 10 Example Result of Wavelength Repeatability Check 5 9 ...
Страница 286: ...5 1 Check with Instrument Main Unit Alone Fig 5 14 Example Result of Noise Level Check 5 12 ...
Страница 289: ...5 1 4 Fig 5 18 Example Result of Baseline Flatness Check 5 15 ...
Страница 292: ...5 1 Check with Instrument Main Unit Alone Fig 5 22 Example Result of Baseline Stability Check 5 18 ...