3.3 Sample-by-sample Measurement (6-cell manual mode)
*
7.
Setting of System Conditions
8.
Setting of Printing Conditions
* For steps 7 and 8, refer to section 3.2.1..
9.
Measurement of Standard Solutions
GUIDE
: When 2) Factor is selected for Curve Type in
step 4, standard solution measurement is not
carried out. So proceed to step 10.
(1) To determine the measuring conditions set in the
preceding steps, press the <9> key (Meas. Screen)
and then [ENTER] key. Or press the [MEAS.
SCREEN] key.
(2) The auto zero sample setting screen (Fig. 3-63)
appears. Set an auto zero sample to cell A.
The lamp is lit for 30 seconds and the current
measured value is indicated so that you can check if
execution of auto zero is necessary.
Upon completion of setting, press the [AUTOZERO]
key. Auto zero will be executed at the position of cell
A. When avoiding auto zero, press the [CLEAR] key
to cancel auto zero.
NOTE
:
For the first measurement under the set
conditions, be sure to execute auto zero.
For details of the auto zero sample, refer to
Explanation 3-10 and Explanation 3-11.
Photometry 600.0 nm 0.000 ABS
CELL A
[Info]
Set autozero sample to Cell A.
AUTOZERO:Execute Autozero; CLEAR:Cancel
Fig. 3-63 Auto Zero Sample Setting Screen
3 - 100
Содержание U-5100
Страница 280: ...5 1 Check with Instrument Main Unit Alone Fig 5 6 Example Result of Wavelength Accuracy Check at 484 6 nm 5 6 ...
Страница 283: ...5 1 2 Fig 5 10 Example Result of Wavelength Repeatability Check 5 9 ...
Страница 286: ...5 1 Check with Instrument Main Unit Alone Fig 5 14 Example Result of Noise Level Check 5 12 ...
Страница 289: ...5 1 4 Fig 5 18 Example Result of Baseline Flatness Check 5 15 ...
Страница 292: ...5 1 Check with Instrument Main Unit Alone Fig 5 22 Example Result of Baseline Stability Check 5 18 ...