4.6.4
4.6.4 How to Return 6-Cell Holder
1.
Removal of Present Cell Holder
Firstly, remove the mask for micro cell when mounted to the
single cell holder. Then, remove the single cell holder or
rectangular long-path cell holder, whichever mounted, by
loosening the screw shown in the figure below.
Screw
Fig. 4-77 Removal of Single Cell Holder
Screw
Fig. 4-78 Removal of Rectangular Long-Path Cell Holder
2.
Mounting of 6-cell Holder
Mount the 6-cell holder, referring to the 6-cell holder
mounting method described in “b.” of 2.4.5.
3.
Change of 6-cell Mode Setting
The 6-cell mode setting needs to be changed.
Turn the 6-cell mode setting from OFF to ON, referring to
section 2.5.4.
4 - 59
Содержание U-5100
Страница 280: ...5 1 Check with Instrument Main Unit Alone Fig 5 6 Example Result of Wavelength Accuracy Check at 484 6 nm 5 6 ...
Страница 283: ...5 1 2 Fig 5 10 Example Result of Wavelength Repeatability Check 5 9 ...
Страница 286: ...5 1 Check with Instrument Main Unit Alone Fig 5 14 Example Result of Noise Level Check 5 12 ...
Страница 289: ...5 1 4 Fig 5 18 Example Result of Baseline Flatness Check 5 15 ...
Страница 292: ...5 1 Check with Instrument Main Unit Alone Fig 5 22 Example Result of Baseline Stability Check 5 18 ...