3.2.3
3 - 61
(3) The Photometry screen (Fig. 3-36) appears.
Photometry
CELL A
1 Meas.PARAM.
Data Mode CONC
2 Sample
No.of WL 1
3 Curve
WL1(nm)
600.0
4 Curve Data
Delay(s) 0
5 6 Cell
6 System
7 Print
8 Save Method
9 Meas.Screen
Select function. : _
Fig. 3-36 Photometry Screen
(4) To set the measuring conditions, press the <1> key
(Meas. PARAM.) and then [ENTER] key. Or select
“Meas. PARAM.” with the [
▲
] or [
▼
] key and press the
[ENTER] key or [
►
] key.
(5) The measuring condition (Meas. PARAM.) screen (Fig.
3-37) appears.
Photometry
CELL A
1 Meas.PARAM.
Data Mode
CONC
2 Sample
No.of WL 1
3 Curve
WL1(nm)
600.0
4 Curve Data
Delay(s) 0
5 System
6 Print
7 Save Method
8 6 Cell
9 Meas.Screen
Select function. : _
□
1
CONC □
2
ABS □
3
%T □
4
DNA
Fig. 3-37 Measuring Condition (Meas. PARAM.)
Screen
(6) For Data Mode, press the <4> key (DNA) and then
[ENTER] key to select DNA measurement.
(7) With the [
▲
] or [
▼
] key, set the number of pairs of
wavelengths (WL pair No.), background correction
(Bkgd. CORR.), wavelength (WL), correction
wavelength (CORR. WL) and initial delay time (Delay).
For details of each parameter, refer to Table 3-27.
(8) After completion of setting, return to the Photometry
screen by pressing the [RETURN] key or [
◄
] key.
Содержание U-5100
Страница 280: ...5 1 Check with Instrument Main Unit Alone Fig 5 6 Example Result of Wavelength Accuracy Check at 484 6 nm 5 6 ...
Страница 283: ...5 1 2 Fig 5 10 Example Result of Wavelength Repeatability Check 5 9 ...
Страница 286: ...5 1 Check with Instrument Main Unit Alone Fig 5 14 Example Result of Noise Level Check 5 12 ...
Страница 289: ...5 1 4 Fig 5 18 Example Result of Baseline Flatness Check 5 15 ...
Страница 292: ...5 1 Check with Instrument Main Unit Alone Fig 5 22 Example Result of Baseline Stability Check 5 18 ...