3.2.4
3 - 93
(5) Return to the WL Scan screen by pressing the
[RETURN] key.
9.
Setting and Measurement of Samples
(1) To determine the measuring conditions set in the
preceding steps, press the <8> key (Meas. Screen)
and then [ENTER] key. Or press the [MEAS.
SCREEN] key.
(2) The guidance on sample setting (Fig. 3-30) appears.
On the screen, displayed from the left are the cell
number and sample name. Cell A is used for baseline
correction. Set a baseline correction sample (blank)
to cell A. And set the corresponding samples to cells
1 to 5. Upon completion of setting, press the [START]
key. After baseline correction with cell A, sample 1
will be measured.
WL Scan
CELL A
[Info]
Set samples to Cell No..
Cell A Blank
Cell 1
Sample1
Cell 2 Sample2
Cell 3 Sample3
Cell 4
-
Cell 5
-
START:Meas.Baseline and Sample1.: _
Fig. 3-58 Guidance on Sample Setting
10. Measurement of Sample Solutions
(1) Upon completion of sample 1 measurement, the
guidance shown in Fig. 3-59 appears. To record data,
perform the save or print operation. To save data,
select the <3> key (Save). To print data, press the
[PRINT] key. Pressing the <1> to <5> keys allows the
processings shown in Table 3-41. For proceeding to
the next measurement, press the [START] key.
Содержание U-5100
Страница 280: ...5 1 Check with Instrument Main Unit Alone Fig 5 6 Example Result of Wavelength Accuracy Check at 484 6 nm 5 6 ...
Страница 283: ...5 1 2 Fig 5 10 Example Result of Wavelength Repeatability Check 5 9 ...
Страница 286: ...5 1 Check with Instrument Main Unit Alone Fig 5 14 Example Result of Noise Level Check 5 12 ...
Страница 289: ...5 1 4 Fig 5 18 Example Result of Baseline Flatness Check 5 15 ...
Страница 292: ...5 1 Check with Instrument Main Unit Alone Fig 5 22 Example Result of Baseline Stability Check 5 18 ...