3.2 Automatic Continuous Measurement (6-cell auto mode)
3 - 84
(2) The 6-cell condition (6 Cell) screen (Fig. 3-52) appears.
To set each item on the 6-cell condition (6 Cell) screen,
press the [ENTER] key. Or select each item with the
[
▲
] or [
▼
] key and make your setting according to the
guidance. For details of each parameter, refer to
Table 3-37.
WL Scan
CELL A
^1 Meas.PARAM.
6 Cell Mode
Auto
2 Sample
No.of Sample 1
3 6 Cell
4 System
5 Peak PARAM.
6 Print
^7 Save Method
8 Meas.Screen
Select function. : _
□
1
Auto □
2
Manual
Fig. 3-52 6-cell Condition (6 Cell) Screen
Table 3-37 “6 Cell” Setting Parameters
Setting Item
Description
6 Cell Mode
Select a 6-cell operation mode for measurement from
the following.
[1] Auto
By setting a baseline sample and unknown samples
to the 6-cell holder, it is possible to automatically turn
the 6-cell holder and automatically measure one
sample after baseline correction.
[2] Manual
Baseline correction and sample measurement are
performed manually. It is required to manually move
the cell and execute baseline correction.
(For details of the Manual mode, refer to section
3.3.4..)
No. of
Sample
Set the number of samples to be measured.
This parameter is settable in a range of 1 to 5.
(3) After setting each item, return to the WL Scan screen by
pressing the [RETURN] key or [
◄
] key.
5.
Setting of System Conditions
(1) To set the system conditions, press the <4> key
(System) and then [ENTER] key. Or select “System”
with the [
▲
] or [
▼
] key and press the [ENTER] key or
[
►
] key.
Содержание U-5100
Страница 280: ...5 1 Check with Instrument Main Unit Alone Fig 5 6 Example Result of Wavelength Accuracy Check at 484 6 nm 5 6 ...
Страница 283: ...5 1 2 Fig 5 10 Example Result of Wavelength Repeatability Check 5 9 ...
Страница 286: ...5 1 Check with Instrument Main Unit Alone Fig 5 14 Example Result of Noise Level Check 5 12 ...
Страница 289: ...5 1 4 Fig 5 18 Example Result of Baseline Flatness Check 5 15 ...
Страница 292: ...5 1 Check with Instrument Main Unit Alone Fig 5 22 Example Result of Baseline Stability Check 5 18 ...