4.1 Loading or Deleting the Saved Data
when the data of sample measurement is included.
If wavelength scan or time scan data is opened, the
calibration curve data cannot be found.
Therefore, you must return to Step 2 “Calibration Curve
Data Loading Procedure” and select the data again.
[When the open data is standard data alone]
Photometry
CELL 1
ID ABS
CONC
STD1 0.001 0.000
STD2 0.102 10.00
STD3 0.205 20.00
Select function. :_
1)Sample Meas. 2)Curve 3)Save
Example of screen showing measured data of
3 standard samples
Fig. 4-14 Standard Data-only Screen
Sample measurement is allowed by selecting the <1>
key (for Sample).
[When the open data includes sample data]
Photometry
CELL 1
ID
ABS
CONC
11 0.001 0.000
12 0.102 10.10
13 0.205 20.05
14 0.172 17.04
15 0.051 5.10
Select function. :_
1)Meas. 2)Curve 3)Save 4)Print
Example of screen showing measured data of
15 unknown samples
Fig. 4-15 Sample Data-included Screen
When you select the <1> key (for Meas.) and then the
<1> key (for New Meas.), the data area of the open file
is cleared and sample measurement is allowed with
the same calibration curve as that of this file.
4 - 12
Содержание U-5100
Страница 280: ...5 1 Check with Instrument Main Unit Alone Fig 5 6 Example Result of Wavelength Accuracy Check at 484 6 nm 5 6 ...
Страница 283: ...5 1 2 Fig 5 10 Example Result of Wavelength Repeatability Check 5 9 ...
Страница 286: ...5 1 Check with Instrument Main Unit Alone Fig 5 14 Example Result of Noise Level Check 5 12 ...
Страница 289: ...5 1 4 Fig 5 18 Example Result of Baseline Flatness Check 5 15 ...
Страница 292: ...5 1 Check with Instrument Main Unit Alone Fig 5 22 Example Result of Baseline Stability Check 5 18 ...