5.2 Check with Pen Type Low-Pressure Mercury Lamp Available at Option
5.2.3 Resolution
(1) Display the Performance Check (Option) screen (Fig. 5-39).
For display method, follow the procedure starting from
section 5.2.. Using the [
▲
]/[
▼
] key, move the cursor to
Resolution Hg (546.1nm), and then select [1] Yes.
Resolution with the Hg lamp (546.1 nm) will be checked. If
check is not desired actually, select [2] No.
Performance
CELL A
(Option)
WL Accuracy Hg (253.7 nm)
WL Accuracy Hg (435.8 nm)
WL Accuracy Hg (546.1 nm)
WL Repeatability Hg
Resolution Hg (546.1 nm)
Hardware
Print Report
(Set Hg Lamp)Execute it? :_
□
1 Yes □
2 No
Fig. 5-39 Performance Check (Option) Screen
(2) The guidance for informing the parameter setting process
(Fig. 5-40) appears, and parameters will be set and
resolution with the Hg lamp (546.1 nm) will be checked.
Setting PARAM....
Fig. 5-40 Guidance for Parameter Setting Process
(3) After the check of resolution with the Hg lamp (546.1 nm),
the screen after this check is displayed as shown in Fig.
5-41. Measurement result is indicated on the Resolution
Hg (546.1nm) line followed by OK when the judgment
standard is met or NG if not met. If NG is indicated, check
the resolution again in the correct procedure. If NG recurs,
refer to section 6.8.
5 - 30
Содержание U-5100
Страница 280: ...5 1 Check with Instrument Main Unit Alone Fig 5 6 Example Result of Wavelength Accuracy Check at 484 6 nm 5 6 ...
Страница 283: ...5 1 2 Fig 5 10 Example Result of Wavelength Repeatability Check 5 9 ...
Страница 286: ...5 1 Check with Instrument Main Unit Alone Fig 5 14 Example Result of Noise Level Check 5 12 ...
Страница 289: ...5 1 4 Fig 5 18 Example Result of Baseline Flatness Check 5 15 ...
Страница 292: ...5 1 Check with Instrument Main Unit Alone Fig 5 22 Example Result of Baseline Stability Check 5 18 ...