4.3.6
(2) For wavelength scan data, the guidance for
wavelength scan (Fig. 4-44) is displayed. And, for
time scan data, the guidance for time scan (Fig. 4-45)
is displayed. On each screen, select the <1> key (for
Trace).
Select function. :_
[1]Trace [2]Peak [3]Smoothing [4]List
Fig. 4-44 Guidance for Wavelength Scan
Select function. :_
[1]Trace [2]Smoothing [3]List
Fig. 4-45 Guidance for Time Scan
3. Tracing
Data
For wavelength scan data, the WL Scan/Trace screen (Fig.
4-46) is displayed. And, for time scan data, the Time
Scan/Trace screen (Fig. 4-47) is displayed. The cursor on
each screen can be moved by pressing the [
◄
]/[
►
] key.
The wavelength and photometric value at the present
cursor position are indicated at the top left of each screen.
Thus, the photometric values at the aimed-at wavelength
and time point can be read.
WL Scan/Trace
CELL A
523.0 nm 0.650 ABS
◄
►
Fig. 4-46 WL Scan/Trace Screen
400.0 525.0 650.0 nm
ABS
0.800
0.000
+