4.1.4
(4) For loading the saved calibration curve, press the <1>
key (for Load/Delete Data) and then the [ENTER] key
or press the [
▲
]/[
▼
] key to select Load/Delete Data
and then the [ENTER] key. The Load/Delete Data
screen (Fig. 4-13) appears.
Load/Delete Data
CELL A
1 NO3-(Hitachi Taro)
PHT 2010/10/10
2 Cr6+(Hitachi Hanako) PHT 2010/03/30
File name
Measurement
mode
Date of file
saving
Select function. :_
[1]Load [2]Delete [3]Check [4]Type [5]Sort
◄
►
( 1/ 1)
Fig. 4-13 Load/Delete Data Screen
(5) On this screen, the saved data files are listed.
The photometric data including the calibration curve is
indicated by PHT in the column of measurement mode.
For checking the parameters according to which the
saved data had been measured, press the <3> key (for
Check) and then the [ENTER] key. The outline of the
selected Method is now displayed. After check, you
can return to the previous display by pressing the
[RETURN] key. Select the data, in which the desired
calibration curve is saved, by the [
▲
]/[
▼
] key and press
the [ENTER] key. The saved data will be displayed.
NOTE :
Please use the data that 6- Cell Mode measured by
turning on without fail for the reading measurement when you
measure 6- Cell Mode in the monitor configuration while turned on.
Please use data to measure 6- Cell Mode by turning off without fail
for the reading measurement when you measure 6- Cell Mode in
the monitor configuration while turned off. Please use Data
measured with Auto Sipper without fail for the reading
measurement when measuring it by using Auto Sipper (another
sales).
3. Sample
Measurement
(1) The contents of data are displayed. When the
contents are only standard data, the list shown in Fig.
4-14 appears, and that shown in Fig. 4-15 appears
4 - 11
Содержание U-5100
Страница 280: ...5 1 Check with Instrument Main Unit Alone Fig 5 6 Example Result of Wavelength Accuracy Check at 484 6 nm 5 6 ...
Страница 283: ...5 1 2 Fig 5 10 Example Result of Wavelength Repeatability Check 5 9 ...
Страница 286: ...5 1 Check with Instrument Main Unit Alone Fig 5 14 Example Result of Noise Level Check 5 12 ...
Страница 289: ...5 1 4 Fig 5 18 Example Result of Baseline Flatness Check 5 15 ...
Страница 292: ...5 1 Check with Instrument Main Unit Alone Fig 5 22 Example Result of Baseline Stability Check 5 18 ...