4.3 Data Processing
4.3.6 Data Tracing (Wavelength/Time Scan Data)
1. Displaying
Data
[For check from saved data]
For check by opening the saved data, open the saved
wavelength scan data or time scan data according to
section 4.1.1..
[During measurement]
When measurement is in progress with WL Scan or Time
Scan selected in Main Menu, you should wait till the end of
the measurement. For stopping the measurement, press
the [STOP] key.
2.
Data Processing Screen
(1) For wavelength scan data, the WL Scan screen (Fig. 4-
42) appears, and the Time Scan screen (Fig. 4-43)
appears for time scan data. Then, select the <1> key
(for Data).
WL Scan
CELL A
ABS
0.800
0.000
400.0 525.0 650.0 nm
START:Meas.; Select function.:_
[1]Data [2]Scale [3]Save [4]Print [5]Baseline
Fig. 4-42 Wavelength Scan Data
Time Scan
CELL A
ABS
0.800
0.000
0 60.0 120 s
START:Meas.; Select function.:_
[1]Data [2]Scale [3]Save [4]Print
Fig. 4-43 Time Scan Data
4 - 38