5.3.1
5.3 Wavelength Calibration
Wavelength calibration is required if the instrument performance
specification is not satisfied in the wavelength accuracy check.
The method of wavelength calibration is explained here.
5.3.1 Wavelength Calibration Method
(1) After startup of the instrument, press the [MAIN MENU] key
to display the Main Menu screen (Fig. 5-43). For selection
of system setup, press the <6> key (System Setup) and
then the [ENTER] key or press the [
▲
]/[
▼
] key to select
System Setup followed by the [ENTER] key.
Main Menu
CELL A
1 Photometry
2 WL Scan
3 Time Scan
4 Method Menu
5 Data Menu
6 System Setup
7 Performance
8 Data Display
Select function. : _
Fig. 5-43 Main Menu Screen
(2) The System Setup screen (Fig. 5-44) appears.
For selecting wavelength calibration, press the <1> key (WL
CALIB.) and then the [ENTER] key or press the [
▲
]/[
▼
] key
to select WL CALIB. followed by the [ENTER] key.
System Setup
CELL A
1 WL CALIB.
2 Lamp Usage
3 Date and Time
4 Printer Setup
5 6 Cell Mode
6 Auto Lamp OFF
7 Language
Select function. : _
Fig. 5-44 System Setup Screen
5 - 33
Содержание U-5100
Страница 280: ...5 1 Check with Instrument Main Unit Alone Fig 5 6 Example Result of Wavelength Accuracy Check at 484 6 nm 5 6 ...
Страница 283: ...5 1 2 Fig 5 10 Example Result of Wavelength Repeatability Check 5 9 ...
Страница 286: ...5 1 Check with Instrument Main Unit Alone Fig 5 14 Example Result of Noise Level Check 5 12 ...
Страница 289: ...5 1 4 Fig 5 18 Example Result of Baseline Flatness Check 5 15 ...
Страница 292: ...5 1 Check with Instrument Main Unit Alone Fig 5 22 Example Result of Baseline Stability Check 5 18 ...