2.14 The Weight Shown on the Scale Display Is Flashing
Chapter 2: Troubleshooting Procedures
2.14 The Weight Shown on the Scale Display Is Flashing
1. The weight on the stretcher is more than 699.9 lb (317.5 kg).
Yes
No
↓
→
Remove the excess weight, and take a weight reading. If this
solves the problem, go to “Final Actions” on page 2-11.
Otherwise, go to step 2.
2.
Look for an interference between the upper and lower frames. There is an
interference.
Yes
No
↓
→
Calibrate the scale (see “Calibration” on page 4-107). If this does
not solve the problem, replace the scale P.C. board (refer to
procedure 4.34). If either of these solve the problem, go to “Final
Actions” on page 2-11. Otherwise, call Technical Support.
3.
Remove the interference, and then take a weight reading. The weight
shows correctly.
Yes
No
↓
→
Call Technical Support.
4.
Go to “Final Actions” on page 2-11.
Page 2 - 26
Hill-Rom® Transport, Procedural, and Specialty Stretchers
Service Manual (144386 REV 2)
Содержание P8000